Page 1 - Cleaningupyour

CleaningupyoursampleprepprocessesJuliaC.Drees,Ph.D.,DABCCScientificDirector—ChemistryKaiserPermanenteRegionalLaboratoriesBerkeleyandRic

Page 2 - Financialdisclosures

Extractionmethods’relativesimplicity/ro bustnessSPE PPT LLESimpler MoreRobustBestWorstAdaptedfromJudyStone,Ph.D.

Page 3 - LearningObjectives

Regardlessofextractionmode,dothis: Addinternalstandard(IS)asFIRSTstepinextraction Evaluateforionsuppression: Qualitative(post‐colu

Page 4 - CommonSamplePrepOptions

Quantitativecharacterizationofyourextraction:recovery,efficiency,matrixeff ects SampleA:Neatsample(exhibits100%recovery,nomatrixeff

Page 5 - Dilution

Automation Onlineextraction Automatedliquid handlers Casehistoriesinautomation

Page 6 - Proteinprecipitation(PPT)

OnlineSPEskipsevaporationandreconstitutionstepsSymbiosisproductliterature,Spark‐Holland

Page 7 - Challengingtoautomate

Onlineex traction‐ mechanicsSchebb,Anal.Methods 2011;3:4206‐portswitchingvalve

Page 8 - SolidPhaseExtraction(SPE)

Considerationsforfeasibilityofonlineextraction Evaluatecycletimesofextraction&LCrunwithcare:Isthroughput adequate? Aretwoextr

Page 9 - Comparingextractionmodes

ConsiderationsforOnlineExtraction‐ AnalyticalIssues # ofsolvents,solventtypesforwash&regenerationofSPEcolumn(solventselectvalve

Page 11 - Regardlessofextraction

AdaptedfromJudyStone,Ph.D.

Page 12

FinancialdisclosuresNothingtodisclose

Page 13 - Automation

AutomatedLiquidHandling1,2,4,8,&/or96pipettingchannelsTubesand96wellplatesCandotubetoplate(startingwithpatientsampletube)

Page 14 - OnlineSPEskipsevaporation

ConsiderationsforbestvaluefromALH Disposabletipsvs.washable(consumablecostvs.carryoverrisk,speedofeach?) #ofpipettechannels(>

Page 15 - Methods 2011;

ConsiderationsforAutomatedLiquidHandling– AnalyticalIssues Minimumandmaximumpipettable volumes(andwithwhatprecision)? Workingwithdiff

Page 16

Auto mationchoices OnlineextractionVendor“complete”solutionsThermo‐Scientific/CohesiveTechnologiesTLXTurbo‐FlowonlineSPEextractionSpark‐H

Page 17 - Extraction‐ AnalyticalIssues

OnlineExtractionvs.AutomatedLiquidHandling(ALH)Parameter Online ALHTech handsontime/sample Less MoreLCsophisticationrequired More LessFlexib

Page 18 - AutomatedLiquidHandling

AutomatedLiquidHandling–CaseHistoriesSampleprepfor25‐OHVitaminDanalysisbyLC‐MSMSThankstomycolleagueswhohelpedtroubleshoottheseca

Page 19

1. Transfer 50 µL IS from reservoir to plate (8 tips, pick up & dispense 12 times)IS reservoir2. Transfer serum from tubes to plate (8x12 tips) &a

Page 20

3. Move plate to 2ndALH (96 channel head) 4. Transfer serum + IS to PPT/phospholipid removal plate5. Transfer precipitating reagent to extraction plat

Page 21

Case1–TheTouch‐OffTaskPipetserumandinternalstandard(IS)toa96‐wellplateDesiredintra‐assayprecision(withinplate)=<5%CV

Page 22 - Handling– AnalyticalIssues

DevelopmentHistoryOriginalProtocol:50µLISdispensedintoplate50µLserumadded,mixwithtipsIntra‐assayprecision4‐6%RevisedProt

Page 23 - Auto mationchoices

LearningObjectivesAfterthispresentationyouwillbeableto: Liststrengthsandweaknessesofcommonsamplepreparationtechniquesusedforclinic

Page 24 - LiquidHandling(ALH)

Problem:%CV↑to6‐9%whenISvolume↓from50µLto25µLRe‐measurebottomofplate dimensionswcalipers&adjustheightofdispense–notfi

Page 25 - CaseHistories

SolutionDispenseserum1stDispenseIS2nd‐ touchofftoserum(liquid)insteadofquestionabletouchoffto(dry)plateIntra‐assay%CVdecreasedt

Page 26 - Tube to

2. Transfer 50 25 µL IS from reservoir to plate (8 tips, pick up & dispense 12 times) (8x12 tips & mix)IS reservoir1. Transfer serum from tube

Page 27 - Stamping

Case2– M issing filtratesTaskPrecipitateserum+ISinahybrid(protein&phospholipidremoval)filtrationplate–transferfiltratetocolle

Page 28 - Case1–TheTouch‐Off

ProblemFirstfewplateslookgoodThenafewwellsineveryplate–nofiltratetransferredtocollectionplateAllotherwe llsinthoseplateshave

Page 29 - DevelopmentHistory

InvestigationChangeofplatelot?– notfixedBreakthroughoffiltrationmembraneorquickflowthroughinsurroundingwellscausinginadequatevacu

Page 30

Eureka!InitialstudiesusedfrozenaliquotsofpooledserumHadrecentlyswitchedtofreshpatientsamplesVendorvalidatedwithfrozenratserumHyp

Page 31 - Solution

Solution:crash,wait,then“sip”&transferCrashinplain,conicalbottom,1mL,plateINSTEADofcrashingINSIDEhybridfiltrationplateWait4

Page 32

3. Move plate to 2ndALH (96 channel head) 5. Wait 4 min for ppt to settle. Transfer supernatant to extraction plate4. Transfer precipitating reagent t

Page 33 - Case2– M issing filtrates

Lessonslearned Givethelowtechsolutionatry Extractionautomationisaboutsurfacechemistry:CharacteristicsofdifferentplasticsSurfacet

Page 34

CommonSamplePrepOptionsDilutionProteinCrashorUltrafiltration(PPT)Liquid‐Liquidextraction(LLE)SolidPhaseextraction(SPE)–canbeoffline

Page 35 - Investigation

Case3–staticclingoftipstopipettehead(tipsnotshucked)

Page 36

SolutionPurc hasetipswithoutplasticfilmwrappingofindividualtipboxe sOnlyre‐usereagenttipsafewtimes(staticbuildsupwithre‐use)Sta

Page 37

InconclusionAtte ntiontotrivialdetail&persistence wasthekeytosuccessAutomationmadepossibleextractionof2,500samples/daybytwot

Page 38 - Plate Stamping:

Thanksforyourattention

Page 39 - Lessonslearned

DilutionUsuallyforurinedrugconfirmationsTypically10‐20xdilutionAddafiltrationstepforadded robustness

Page 40

Proteinprecipitation(PPT)Avoidaddinginternalstandard(IS)toprecipitationsolventIdeallyaratioof3:1precipitant:sampleOptionstoinclude

Page 41

Liquid‐LiquidExtraction(LLE)BestforneutralanalytesIdeally10:1solvent:sampleMixtureofpolarandnonpolarsolventsMorepolar→morejunkPha

Page 42 - Inconclusion

SolidPhaseExtraction(SPE)MostchancesforusingscientificapproachforoptimizationIonexchangeSPEismostselective(chargedanalytes)Washag

Page 43 - Thanksforyourattention

ComparingextractionmodesRankordered 1=best,4=worst.*=IonexchangeSPEmodeetc.Differentiator SPE PPT LLE OnlineCharged Anal